Sandhill Frequently Asked Quetions
For the ZepHr™ & Sleuth™ Systems, and Computers

1. I am trying to format the compact flash card and the computer flashes a quick message, "completed format". I did not see the usual black popup window to verify deletion of files and format card. Is this OK?
2. When I try to format the compact flash I get an error, "Please insert compact flash into drive", but a card is already in the recorder. What should I do?
3. I am unable to calibrate the catheter for the ZepHr/Sleuth.
4. I am unable to find the proper ZepHr/Sleuth catheter protocol that I want to use in the list
5. I am using the ZepHr/Sleuth with LES Locator. How do I select display pressure and how far is the port from the tip?
6. I need to change the date and time on the ZepHr/Sleuth.
7. My datalogger says, "Lead Off". What should I do?
8. The datalogger says, "Low battery". What should I do?


1. I am trying to format the compact flash card and the computer flashes a quick message, "completed format". I did not see the usual black popup window to verify deletion of files and format card. Is this OK?
• You are not logged on as the administrator. You must see the black pop up and answer the 2 questions to format properly. Log on properly.

back to the top


2. When I try to format the compact flash I get an error, "Please insert compact flash into drive", but a card is already in the recorder. What should I do?
• Confirm you logged on the computer with administrative privileges. If not, log on properly.
• Is the correct drive selected under the options button? If not, select correct drive.
• Is there a green light on the reader illuminated? If not, reboot the computer.
• You must use an official Sandhill Compact Flash Card. If not, change to a Sandhill tested and approved Compact Flash Card, Part# Z00-1200-256.

back to the top


3. I am unable to calibrate the catheter for the ZepHr/Sleuth.
• You must verify proper catheter connection; verify protocol # on display matches package reorder #.
• Verify skin was prepped w/ alcohol prep pad before applying electrode patch. Prep finger to be in buffer as well.
• Verify external electrode is connected to patient and the patient's finger is in the buffer.
• Verify that all pH and impedance sensors are completely submersed in the buffer.

back to the top


4. I am unable to find the proper ZepHr/Sleuth catheter protocol that I want to use in the list.
• Go to the ZepHr/Sleuth "start new patient window" and click onthe option button in lower left corner.
• Go to edit protocol list and select the proper protocol to be displayed.
• If this does not work, please call Sandhill Support 1-800.468.4556

back to the top


5. I am using the ZepHr/Sleuth with LES Locator. How do I select display pressure and how far is the port from the tip?
• Open the ZepHr/Sleuth program and go into the options in lower left corner.
• Select "show pressure".
• The Locator port is 6cm above the distal esophageal pH sensor.

back to the top


6. I need to change the date and time on the ZepHr/Sleuth.
• You must REMOVE the Compact Flash Card in the ZepHr/Sleuth before starting this procedure.
• Insert batteries and push the light bulb button.
• Then push the "1" button. Arrow over to next and press enter ("1")button, select "setup" and change the time.

back to the top


7. My datalogger says, "Lead Off". What should I do?
• Take a few sips of water and wait a few minutes
• Check and make sure the probe is connected tightly. You may try to disconnect and reconnect it.
• Take a few more sips of water and wait a few minutes.
• If the problem continues, call your clinician

back to the top


8. The datalogger says, "Low battery". What should I do?
• Open the battery compartment and replace with the appropriate alkaline battery.

back to the top